Is it conceivable that the control was run once because the key result came from the same run? I can see a reviewer asking for it in all three figures, whereas they may drafted it only in one

gus_massa 2 days ago

The horizontal label is fine, it says Pon S in all images. (I guess a wrong label would be obvious to detect for specialists.)

The problem are the vertical labels

In Figure 1e it says: "MT1+2", "MT2" and "MT1"

In Figure 3a it says: "5'-CR1", "CR2" and "3'-UTR"

In Figure 3b it says: "CR2", "CR3" and "CR4"

  • [removed] 2 days ago
    [deleted]
shpongled 2 days ago

Based on the images, it is inconceivable that these are from the same run (see the dramatically different levels of TRF-S in each gel. One column/lane = one sample). This isn't something that would be included because of a reviewer - loading controls are required to meaningfully interpret the results (e.g. the data is useless without such a control).